Sequencing Protocol
Sequencing Protocol

What is the sequencing protocol?

The following protocol utilises ABI BigDye terminator v3.1 and ABI3730xl DNA analyser.

1. Sequencing Reaction Set-up:

Sequencing reactions can be performed as follows:

Please perform your sequencing reactions in 200 microliter centrifuge tubes.

Component

Volume

Template DNA (300-600 ng)

x µL

Primer (3.2 uM)

1 µL

ABI BigDye terminator v3.1*

1 µL

5x Sequencing Buffer

3.5 µL

Milli-Q water

y µL

Total volume

20 µL

* The volume of terminator used is 1/8 reaction dilution.

2. Sequencing Cycling Conditions:

Please use the following cycling conditions with a PE Applied Biosystems GeneAmp PCR System 9700 thermal cycler. If you have other thermal cycler, please consult Applied Biosystems (www.appliedbiosystems.com) for cycling conditions.

  • 96 ° C 1 min
  • 96 ° C 10s
  • 50 ° C 5s
  • 60 ° C 4 min
  • 25 cycles
  • 4 ° C (infinity)

3. Sample Submission

After cycling reactions, seal the centrifuge tubes (containing 20 µL of reaction volume) with parafilm or equivalent tape. Mail your samples together with a copy of the order details to AgGenomics Pty. Ltd., Victorian AgriBiosciences Centre, 1 Park Drive, Bundoora, Victoria 3086, Australia.

For sequencing protocol discussions, please call us on  (03) 9479 3931 or email xpress@aggenomics.com.au